• k8凯发

    Scanning electron microscopy protocol 

    Rene Flores Clavo, Francisco Breno S. Teófilo

    Published: 2022-11-10 DOI: 10.17504/protocols.io.j8nlkk796l5r/v1
    AI 解读
    浏览 1406

    Abstract

    This protocol briefly summarizes the basic steps of a scanning electron microscopy processing. The methods adopted for fixation, post-fixation, dehydration, drying in a critical point chamber, sputter coating, and the visualization and acquisition of images are described here.

    Steps

    MATERIAL SELECTION

    1.

    1 mm2 samples were selected in the colony.

    FIXATION

    2.

    Samples were fixed in a solution of 2.5% volume and 0.1Molarity (M) buffer 7.3) , at Room temperature, for 4h 0m 0s.

    WASHING AFTER FIXATION

    3.

    Samples was then rinsed three times in 0.1Molarity (M) (7.3), at Room temperature , for 0h 20m 0s for each rinse.

    POST-FIXATION

    4.

    Samples was then post-fixed with 1% volume osmium tetroxide in 0.1Molarity (M) buffer, 7.3, at Room temperature, for 1h 0m 0s, protected from ambient light.

    WASHING AFTER POST-FIXATION

    5.

    Samples was then rinsed three times in sodium 0.1Molarity (M) buffer 7.3, at Room temperature, for 0h 20m 0s for each rinse.

    DEHYDRATION

    6.

    Samples were dehydrated with ascending acetone series, for 0h 20m 0s minutes each step - 30% volume; 50% volume, 70% volume, 90% volume, 100% volume (last one step, for three times).

    CRITICAL POINT DRYING

    7.

    Samples was then dried using the critical point method with CO2.

    MOUTING ON THE STUBS

    8.

    Samples was then placed on aluminum stubs.

    SPUTTER-COATER

    9.

    Samples was then coated with a layer of 30−40 nm gold using a Balzers SCD 050 sputter-coater.

    OBSERVATIONS AND IMAGE ACQUISITION

    10.

    Observations and photomicrograph aquisitions were obtained using a JEOL JSM 5800LV at 10 kV with SemAfore 5.21 software.