    Real time-quaking induced conversion assay (RT-QUIC)

    Arpine Sokratian, andrew.west west

    Published: 2024-01-15 DOI: 10.17504/protocols.io.6qpvr67kpvmk/v1
    AI 解读
    浏览 1457

    Abstract

    Seeded-amplification assay (SAA) method for the detection of pathology-associated aggregation-templating competent forms of alpha-synuclein and other interacting factors.

    Attachments

    RT-QUIC outline.xlsx

    Steps

    Plate preparation

    1.

    Thaw down a-syn monomer and sonicated fibril aliquots On ice, do not generate bubbles by pipetting

    2.

    Measure monomer concentration via Nanodrop

    Add 3µL of 10x diluted aliquot in PBS onto nanodrop piedestal;

    Parameters: other proteins; coefficient extinction: 5.98; MW: 14.4 kDA

    Perform two measurements and confirm <10% standard error between two measurements

    If necessary, prepare 20X and 30X dilutions to confirm findings.

    Equipment

    ValueLabel
    NanoDrop™ One/OneC Microvolume UV-Vis SpectrophotometerNAME
    UV-Vis SpectrophotometerTYPE
    Thermo ScientificBRAND
    ND-ONE-WSKU
    http://www.thermofisher.com/us/en/home.htmlLINK
    3.

    Calculate the volume of monomer need following: 0.3mg/mL monomer in a reaction mix X 30µLper well

    _30 ul (reaction volume) * 3 (replicates) * ____ (different conditions) = ____ul;

    Dilute the monomer preparation with to 2.5mg/mL in order to filter the protein preparation for a high efficiency

    4.

    Use to filter out possible aggregates or HMW contaminates

    5.

    Meantime measure concentration of sonicated fibrils: prepare serial dilutions: 2x; 4x; 8x in3Molarity (M) Guanidine HCL, volume: 10µL

    Add 3µL onto nanodrop piedestal;

    Parameters: other proteins; coefficient extinction: 5.98; MW: 14.4 kDA

    Blank against 3Molarity (M) Guanidine HCL;

    Perform two measurements for each dilution and confirm <10% standard error between two measurements

    Equipment

    ValueLabel
    NanoDrop™ One UV-Vis SpectrophotometerNAME
    spectrophotometerTYPE
    Thermo ScientificBRAND
    ND-ONE-WSKU
    Sample Volume (Metric): Minimum 1µL; Spectral Bandwidth: ≤1.8 nm (FWHM at Hg 254 nm); System Requirements: Windows™ 8.1 and 10, 64 bit; Voltage: 12 V (DC); Wavelength Range: 190–850 nmSPECIFICATIONS
    6.

    Measure DLS data for sonicated fibrils

    Dynamic Light Scattering measurements

    7.

    Place single zirconia bead into each well of the

    8.

    Prepare serial dilutions of sonicated fibrils:

    Prepare solution: 1mg/mLmeasured PFFs (20µL volume);

    1. 10 ug/ml: 10 ul PFFs + 990 ul of PBS;

    2. 1000 ng/ml: 10 ul PFFs + 90 ul of PBS;

    3. 100 ng/ml: 10 ul PFFs + 90 ul of PBS;

    4. 10 ng/ml: 10 ul PFFs + 90 ul of PBS;

    5. 1 ng/ml: 10 ul PFFs + 90 ul of PBS;

    6. 100 pg/ml: 10 ul PFFs + 90 ul of PBS;

    7. 10 pg/ml: 10 ul PFFs + 90 ul of PBS;

    8. 1 pg/ml: 10 ul PFFs + 90 ul of PBS;

    9. 100 fg/ml: 10 ul PFFs + 90 ul of PBS;

    10. 10 fg/ml: 10 ul PFFs + 90 ul of PBS;

    9.

    Prepare reaction mix:

    0.3mg/mL monomer in a reaction mix + 10micromolar (µM) in PBS (account for 10% volume to add) = ____ 30µL reaction * triplicates * sample number (include standard curve of PFFs serial dilutions, 10 reactions in triplicates)

    [total volume * 0.1% = total volume - (total volume * 0.1%)] - [volume of filtered monomer at __ mg/mL after filter] - [ThT volume] = __ volume of PBS

    10.

    Prepare a standard curve using PFFs dilutions from

    1. 100 ng/ml: 10 ul of #2 (from step 8) + 90 ul of reaction mix;

    2. 10 ng/ml: 10 ul of #3 (from step 8) + 90 ul of reaction mix;

    3. 1 ng/ml: 10 ul of #4 (from step 8) + 90 ul of reaction mix;

    4. 100 pg/ml: 10 ul of #5 (from step 8) + 90 ul of reaction mix;

    5. 10 pg/ml: 10 ul of #6 (from step 8) + 90 ul of reaction mix;

    6. 1 pg/ml: 10 ul of #7 (from step 8) + 90 ul of reaction mix;

    7. 100 fg/ml: 10 ul of #8 (from step 8) + 90 ul of reaction mix;

    8. 10 fg/ml: 10 ul of #9 (from step 8) + 90 ul of reaction mix;

    9. 1 fg/ml: 10 ul of #10 (from step 8) + 90 ul of reaction mix;

    10. PBS: 10 ul of PBS + 90 ul of reaction mix;

    11.

    Transfer reaction to the plate: start filling the plate with a standard curve samples (seal the portion with a sealing foil)

    Proceed the experimental samples and cover the whole plate with a sealing foil. Spin down the plate 500x g,0h 0m 0s

    for 0h 0m 30s

    Equipment

    ValueLabel
    Eppendorf™ 022620572NAME
    Microcentrifuges with a plate rotorTYPE
    EppendorfBRAND
    13-690-003SKU
    http://www.fishersci.com/us/en/home.htmlLINK

    Example of the plate layout:

    12.

    Set up the program on a plate reader

    Equipment

    ValueLabel
    Omega - RT-QuIC / PRION Version - Fluorescence BaseNAME
    BMG Fluostar plate readerTYPE
    BMGBRAND
    0415-102PSKU
    http://www.bmglabtech.com/clariostar-plus/?utm_term=clariostar&utm_campaign=usa.roi.microplatereaders+products&utm_source=adwords&utm_medium&gclid=Cj0KCQjw4ImEBhDFARIsAGOTMj9ZLSjhHBhKWAli17cxj6V54U96GSg3qsfuFrfP1cETTLgLbn-kdUIaAkovEALw_wcBLINK
    13.

    Finish the experiment once standard curve reached the plateau

    EXAMPLE: Plateau is reached at7 hours.

    Data analysis

    14.

    Convert collected data from plate reader into .xlsx type

    EXAMPLE:

    15.

    The main advantage of qQuiC analysis is capability to calculate exact amount of fibril forming units in pathological samples relatively to an appropriate standard curve. The approach of calculation FFUs consists of using standard curve with a range of serial dilutions of evaluated ssFibrils spiked in original matrix of the sample. Obtained ThT FL units for each time point and dilution are needed to define an appropriate threshold value needed to define CTvalues. To find a threshold value we built a program that can find an increase in the rates of changes greater than X present for fluorescence values in each replicate. The program looks at three points at a time and compares the rate of change from the first point to the second point against the second point to the third point. Once it finds the point where the rate of change increases by X present from one jump to the next the program stores that value to calculate the threshold value for each sample at X percent as the cutoff difference in the rates of change. The code loops through all of the replicates >30 times to check every percent threshold for X from 10% to 30% and returns the threshold value that yield the highest R-squared in a standard curve that is made from the mean CT values for each replicate. When threshold value is found CTvalue for each dilution can be determined as a time when ThT fluorescence reaches to a threshold. If R-square is higher than 0.8 than efficacy of the reaction is enough to be able to calculate FFUs. To define FFUs of experimental samples we applied an extracted equation from linear regression curve. Code for a program http://github.com/west-lab/FibrilOptimization/blob/master/FibrilPaperCode_main.py

    This step is required to QC the reaction by evaluating the standard curve and finding the Ct values for experimental conditions

    16.

    Use mean half-time of the reaction by analyzing the Ct values of the standard curve

    17.

    Extract RFU values for each condition

    EXAMPLE:

    18.

    Use the extracted data to generate a plot or a group analysis via GraphPad